Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Stryker Orthopaedics Trident Constrained Liner Inserter/Impactor Tip Recalled by Stryker Howmedica Osteonics Corp. Due to Impactors do not meet the required sterility assurance...

Date: June 3, 2016
Company: Stryker Howmedica Osteonics Corp.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Stryker Howmedica Osteonics Corp. directly.

Affected Products

Stryker Orthopaedics Trident Constrained Liner Inserter/Impactor Tip, 22 mm, 28 mm, and 32 mm. Non Sterile Catalog Numbers 2199-2022, 2199-2028 and 2199-2032. For use in total hip arthroplasty.

Quantity: 1,864 units

Why Was This Recalled?

Impactors do not meet the required sterility assurance level (SAL) of 10-6 when using the recommended steam sterilization method.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Stryker Howmedica Osteonics Corp.

Stryker Howmedica Osteonics Corp. has 87 total recalls tracked by RecallDetector.