Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Stryker Howmedica Osteonics Recalled by Stryker Howmedica Osteonics Corp. Due to During the manufacturing of the affected lot a...

Date: August 6, 2012
Company: Stryker Howmedica Osteonics Corp.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Stryker Howmedica Osteonics Corp. directly.

Affected Products

Stryker Howmedica Osteonics; Stryker T2 Humerus System Teflon Tube, sterile; REF 1806-0073S The T2 Nailing System is designed to treat complex fractures in Humeral, Femoral, Tibial and Ankle applications. The Teflon Tube is an optional instrument used to exchange the Ball Tip Guide Wire with the Smooth Tip Guide Wire once reaming is completed. Stryker Stryker Trauma GmbH 24232 Schonkirchen, Germany distributed in USA by: Howmedica Osteonics Corp., 235 Corporate Drive,

Quantity: 100 units (73 units US; 27 units OUS)

Why Was This Recalled?

During the manufacturing of the affected lot a manufacturing error has led to embrittlement of the tube material of the T2 Humerus Teflon Tube.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Stryker Howmedica Osteonics Corp.

Stryker Howmedica Osteonics Corp. has 87 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report