Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Howmedica Osteonics Corp. Stryker Orthopaedics Specialty Triathlon Left and Right Recalled by Stryker Howmedica Osteonics Corp. Due to The standard Triathlon Femoral Stylus does not fit...

Date: January 24, 2013
Company: Stryker Howmedica Osteonics Corp.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Stryker Howmedica Osteonics Corp. directly.

Affected Products

Howmedica Osteonics Corp. Stryker Orthopaedics Specialty Triathlon Left and Right A//P Sizer Adjustment Housing PER File K2401. For use in knee prosthesis implantation. NON-STERILE

Quantity: 4 units (total of 2 sets)

Why Was This Recalled?

The standard Triathlon Femoral Stylus does not fit into the Specialty Sizers and the "R" and "L" markings on the Right Sizer are reversed.

Where Was This Sold?

This product was distributed to 2 states: DE, NJ

Affected (2 states)Not affected

About Stryker Howmedica Osteonics Corp.

Stryker Howmedica Osteonics Corp. has 87 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report