Shaver Handpiece Sterilization Tray. Model number 272-700-000. For sterilization of Recalled by Stryker Endoscopy Due to STERRAD 100S parameters provided in the Shaver Handpiece...

Date: May 19, 2014
Company: Stryker Endoscopy
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Stryker Endoscopy directly.

Affected Products

Shaver Handpiece Sterilization Tray. Model number 272-700-000. For sterilization of Stryker Endoscopy arthroscopic shaver.

Quantity: 1992 trays

Why Was This Recalled?

STERRAD 100S parameters provided in the Shaver Handpiece reprocessing guide (P10299 Revision A) did not fully consider all worst-case scenarios during validation testing.

Where Was This Sold?

Worldwide Distribution.

About Stryker Endoscopy

Stryker Endoscopy has 238 total recalls tracked by RecallDetector.

Related Recalls