Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

100V NEPTUNE 3 ROVER JAPANESE (0703-003-000) 120V NEPTUNE S ROVER Recalled by Stryker Corporation Due to Bolts on the vacuum pump filter housing and...

Date: December 3, 2024
Company: Stryker Corporation
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Stryker Corporation directly.

Affected Products

100V NEPTUNE 3 ROVER JAPANESE (0703-003-000) 120V NEPTUNE S ROVER (0711-001-000) 230V NEPTUNE 3 ROVER ENGLISH (0703-002-000ENG) 230V NEPTUNE 3 ROVER SPANISH (0703-002-000ES)

Quantity: 14 (Expanded to 21 units on 10/9/25)

Why Was This Recalled?

Bolts on the vacuum pump filter housing and the flange plate were not properly tightened, impacting the function of the Neptune rovers. Systems with loose bolts may fail to provide sufficient suction force or loss of suction during a procedure.

Where Was This Sold?

This product was distributed to 3 states: GA, IL, WA

Affected (3 states)Not affected

About Stryker Corporation

Stryker Corporation has 108 total recalls tracked by RecallDetector.

Related Recalls