Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Berchtold Chromophare Ceiling Mounted Surgical Light System Recalled by Stryker Communications Due to The firm has become aware that there is...

Date: October 11, 2019
Company: Stryker Communications
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Stryker Communications directly.

Affected Products

Berchtold Chromophare Ceiling Mounted Surgical Light System, Catalog #CH13000000

Quantity: 1415 Surgical Lights

Why Was This Recalled?

The firm has become aware that there is a likelihood that a missing snap ring or spanner nut may result in an insufficient mount force that causes the surgical light to fall. If the suspension of the surgical light falls, it may potentially cause serious injuries to Health Care Providers and/or patients.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Stryker Communications

Stryker Communications has 18 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report