Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Cyto-Chex BCT - blood specimen collection device - Immunophenotyping Preservative. Recalled by Streck Due to Complaints of the 2.0ml tube stopper coming loose...

Date: February 23, 2017
Company: Streck
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Streck directly.

Affected Products

Cyto-Chex BCT - blood specimen collection device - Immunophenotyping Preservative. IVD. Sterile. 2.0ml only. Packaged 6-tube pack REF 213559 and 100-tube box REF 218980. Streck.

Quantity: 128,692 tubes

Why Was This Recalled?

Complaints of the 2.0ml tube stopper coming loose during phlebotomy when small diameter tube adapters were not utilized. Such adapters aid with centering small diameter tubes to properly pierce the middle of the stoppers.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Streck

Streck has 1 total recall tracked by RecallDetector.

Related Recalls