Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Straumann Drill 1 Recalled by Straumann Manufacturing, Inc. Due to Drills cannot be inserted into handpiece due to...

Date: November 23, 2016
Company: Straumann Manufacturing, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Straumann Manufacturing, Inc. directly.

Affected Products

Straumann Drill 1, short, ¿ 2.2mm, l 33.0mm, SS-Non sterile. Dental implant accessory Article Number: 044.210 Dental implants drill accessory is intended to prepare osteotomy site for surgical dental implant placement. Dental implants are intended for immediate placement and function on single-tooth and/or multiple tooth applications when good primary stability is achieved, with appropriate occlusal loading, in order to restore chewing function

Quantity: 165 units

Why Was This Recalled?

Drills cannot be inserted into handpiece due to tolerance specification not met

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Straumann Manufacturing, Inc.

Straumann Manufacturing, Inc. has 11 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report