Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Vis-U-All High Temp 5"x15" Self Seal Pouch 200 pouches per box Recalled by Steris Corporation Due to Purple ink located on the outside label of...

Date: December 2, 2020
Company: Steris Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Steris Corporation directly.

Affected Products

Vis-U-All High Temp 5"x15" Self Seal Pouch 200 pouches per box; 4 boxes per case Model Number: 884515

Quantity: 7 cases

Why Was This Recalled?

Purple ink located on the outside label of Vis-U-All Pouches may migrate ( bleed ) following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Steris Corporation

Steris Corporation has 109 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report