Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

The SYSTEM 1E Liquid Sterilant Processing System. The SYSTEM 1E Recalled by Steris Corporation Due to The firm initiated a recall after complaint investigations...

Date: August 1, 2012
Company: Steris Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Steris Corporation directly.

Affected Products

The SYSTEM 1E Liquid Sterilant Processing System. The SYSTEM 1E Liquid Sterilant Processing System is intended for the liquid chemical sterilization of manually cleaned immersible and reusable critical and semi-critical heat sensitive medical devices, including endoscopes and their accessories.

Quantity: 6,051 units

Why Was This Recalled?

The firm initiated a recall after complaint investigations revealed that when the units are left on and not in use at night and/or on weekends the hose disconnects/separates resulting in water leakage that can cause damage to property.

Where Was This Sold?

This product was distributed nationwide across all 50 states.