Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

The 5085 is a general surgical table with high patient weight capacity Recalled by Steris Corporation Due to The 5085 and 5085SRT stainless steel column shrouds...

Date: October 10, 2014
Company: Steris Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Steris Corporation directly.

Affected Products

The 5085 is a general surgical table with high patient weight capacity, extended width capability, and lower minimal table top elevation.

Quantity: 999

Why Was This Recalled?

The 5085 and 5085SRT stainless steel column shrouds may become damaged as a result of misalignment during extreme table articulation, excessive side force on the table column, or pressure on the base of the table. Severe damage to the shrouds may prevent the user from raising or lowering the surgical table.

Where Was This Sold?

This product was distributed to 36 states: AL, AZ, AR, CA, CO, FL, GA, HI, ID, IL, IN, IA, KY, LA, ME, MD, MI, MN, MS, MO, NV, NY, NC, ND, OH, OK, OR, RI, SC, SD, TN, TX, VA, WV, WI, DC

Affected (36 states)Not affected

About Steris Corporation

Steris Corporation has 109 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report