Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

SYSTEM 1E Liquid Chemical Sterilant Processing System Recalled by Steris Corporation Due to Customers indicated fluctuations in environmental conditions sites and...

Date: November 21, 2013
Company: Steris Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Steris Corporation directly.

Affected Products

SYSTEM 1E Liquid Chemical Sterilant Processing System, Distributed by STERIS Corporation, Mentor, OH The SYSTEM 1E Liquid Chemical Sterilant Processing System is intended for the liquid chemical sterilization of manually cleaned immersible and reusable critical and semi-critical heat sensitive medical devices, including endoscopes and their accessories.

Quantity: 6,028 units

Why Was This Recalled?

Customers indicated fluctuations in environmental conditions sites and gradual degradation/deformation of check valve o-rings following frequent contact with S40 sterilant use solution result in nuisance aborts.

Where Was This Sold?

This product was distributed nationwide across all 50 states.