Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Harmony Equipment Management System Recalled by Steris Corporation Due to Dr¿ger Medical GmbH has notified STERIS of the...

Date: August 15, 2013
Company: Steris Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Steris Corporation directly.

Affected Products

Harmony Equipment Management System, Model #603X/80XX, STERIS Corporation, Mentor, OH Harmony Equipment Management Systems (EMS) are used in hospital facilities to provide ready access to medical gasses, electrical power and audio/visual data services, in addition to managing surgical support equipment such as endoscopic equipment, cautery devices and insufflators.

Quantity: 127 units

Why Was This Recalled?

Dr¿ger Medical GmbH has notified STERIS of the potential for a break in the drive screw contained in the motorized ceiling supply unit, a component of the Harmony EMS system. The ceiling supply unit controls the motorized adjustment of the height and rotation of the supply column. Should a break occur in the drive screw, there is potential for the adjustable arm to drop down, causing the EMS sys

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Steris Corporation

Steris Corporation has 109 total recalls tracked by RecallDetector.

Related Recalls