Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

AMSCO C and AMSCO 400 Steam Sterilizers designed for sterilization Recalled by Steris Corporation Due to The control boards alarm set points and temperature...

Date: February 12, 2013
Company: Steris Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Steris Corporation directly.

Affected Products

AMSCO C and AMSCO 400 Steam Sterilizers designed for sterilization of heat and moisture-stable materials used in healthcare facilities.

Quantity: 395 Units

Why Was This Recalled?

The control boards alarm set points and temperature tolerances allowed for erroneous activation of the AD (analog/digital) alarm as temperature fluctuated during operation. When this alarm occurs, the sterilizer locks up and cannot be used until serviced, causing Customer nuisance.

Where Was This Sold?

This product was distributed to 45 states: AL, AK, AZ, AR, CA, CO, CT, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, NE, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, DC

Affected (45 states)Not affected

About Steris Corporation

Steris Corporation has 109 total recalls tracked by RecallDetector.

Related Recalls