Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

SteriGear Urinary Drain Bag with Fig Leaf Cover Recalled by SteriGear, LLC Due to SteriGear notifed customers of circumstances in which The...

Date: January 11, 2011
Company: SteriGear, LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact SteriGear, LLC directly.

Affected Products

SteriGear Urinary Drain Bag with Fig Leaf Cover, 2000 ml, REF 10270, The Fig Leaf, SteriGear 362 S. University Avenue, provo, UT 84601, 1-800-398-3259, SteriGear.com, Sterile, EO Urinary Drainage Collection Kit, for Indwelling Catheter. This device is intended to collect the Urine drained from an indwelling catheter

Quantity: 110,560

Why Was This Recalled?

SteriGear notifed customers of circumstances in which The Fig Leaf urinary drain bag valve can close creating a negative pressure that can suck the non-return valve closed.

Where Was This Sold?

This product was distributed nationwide across all 50 states.