Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

STAT-Site M BHB Test Strips Recalled by Stanbio Laboratory, LP Due to Deterioration in the stability of the Test Strips...

Date: October 2, 2023
Company: Stanbio Laboratory, LP
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Stanbio Laboratory, LP directly.

Affected Products

STAT-Site M BHB Test Strips, used with STAT-SITE M Beta-HB Meter as an in vitro diagnostic for blood ketone values.

Quantity: 2,351

Why Was This Recalled?

Deterioration in the stability of the Test Strips results in diminished low-range detection capability and inability to meet linearity performance claims. The root cause of the stability deterioration in is identified as the change in material that caused a pH shift in the coating solution resulting in precipitation during coating runs. The device defect can cause false negative results that may lead to a delay in treatment of patients with associated conditions such as diabetic ketoacidosis.

Where Was This Sold?

This product was distributed nationwide across all 50 states.