Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

PM1214 NUANCE SR RF Recalled by St Jude Medical Inc. Due to New pacemaker firmware was developed to further mitigate...

Date: August 28, 2017
Company: St Jude Medical Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact St Jude Medical Inc. directly.

Affected Products

PM1214 NUANCE SR RF; PM1230 and PM2230 NUANCE MRI; PM2214 NUANCE DR RF These low voltage (LV) devices are implantable pacemaker pulse generators, intended to be permanently implanted in the body, that have a power supply and electronic circuits that produce a periodic electrical pulse to stimulate the heart. These devices are used as a substitute for the heart's intrinsic pacing system to correct both intermittent and continuous cardiac rhythm disorders.

Quantity: 2741 units

Why Was This Recalled?

New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications.

Where Was This Sold?

Worldwide Distribution