Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Merlin PCS Programmer Software: 3330 The Merlin PCS programmer (model Recalled by St Jude Medical Inc. Due to The firm is advising physicians that exposure to...

Date: April 16, 2018
Company: St Jude Medical Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact St Jude Medical Inc. directly.

Affected Products

Merlin PCS Programmer Software: 3330 The Merlin PCS programmer (model 3650) is a portable, external device used to support the implanted device, and for this recall the software that resides on the programmer (software model 3330) is the method of delivery of the new firmware.

Quantity: 0

Why Was This Recalled?

The firm is advising physicians that exposure to sub-freezing temperatures during the supply chain process caused a transient battery voltage drop for a small number of Confirm Rx Model DM3500 Insertable Cardiac Monitoring (ICM) devices. This drop is an expected and normal behavior for this battery chemistry, but causes an incorrect display of a low battery indicator even after the battery voltage returns to normal.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected