Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

SpineFrontier S-Lift Lock Shaft Inserter Recalled by SpineFrontier, Inc. Due to It has been verified through field use that...

Date: December 23, 2011
Company: SpineFrontier, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact SpineFrontier, Inc. directly.

Affected Products

SpineFrontier S-Lift Lock Shaft Inserter, an instrument used with S-LIFT Intervertebral Body Fusion Device in spinal fusion surgery. Product Usage: The S-Lift Lock Shaft SI50001 mates with SI50002, Implant Inserter, to aid in securing the implant to the whole implant inserter assembly, SI50000.

Quantity: 30 devices

Why Was This Recalled?

It has been verified through field use that there are two possible breakages that may occur when using SI50001 Revision D Lock Shaft Inserter, possibly resulting in instrument fragments entering the sterile field: Impact Cap: The impact cap on the Lock Shaft Inserter can shear off under high-impact force during insertion and removal. Pin: The pin used to guide the Lock Shaft Inserter into the ma

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About SpineFrontier, Inc.

SpineFrontier, Inc. has 33 total recalls tracked by RecallDetector.