Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Quantum Perfusion Systems for CPB Recalled by Spectrum Medical Ltd. Due to The ratchet and pawl had an out of...

Date: March 13, 2023
Company: Spectrum Medical Ltd.
Status: Completed
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Spectrum Medical Ltd. directly.

Affected Products

Quantum Perfusion Systems for CPB, Roller Pumps use peristaltic action to propel blood through the venous and arterial circuit tubing. Catalog Numbers: a) 51-000013-00, b) 51-000008-00

Quantity: 172 units

Why Was This Recalled?

The ratchet and pawl had an out of tolerance geometry which exhibited as the left bobbin would not lock.

Where Was This Sold?

This product was distributed nationwide across all 50 states.