Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

MiniScope Hst ROC Crown 15 Patient Kit Recalled by Specialty Appliance Works, Inc. Due to The Universal Nut, a component of the orthodontic...

Date: March 9, 2015
Company: Specialty Appliance Works, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Specialty Appliance Works, Inc. directly.

Affected Products

MiniScope Hst ROC Crown 15 Patient Kit, and MiniScope Herbst Crown 15 Patient Kit A Herbst Appliance is used to affect changes in the skeletal relationship and/or occlusion of a patient. the Universal Nut is a component of a Herbst Appliance.

Quantity: 14,428 Universal Nuts

Why Was This Recalled?

The Universal Nut, a component of the orthodontic Herbst Appliance was made with 416 SS and not the specified 303 SS.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Specialty Appliance Works, Inc.

Specialty Appliance Works, Inc. has 10 total recalls tracked by RecallDetector.