Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Ultraview SL Command Modules Recalled by Spacelabs Healthcare Inc Due to Modules resetting. This failure mode happens when the...

Date: April 21, 2015
Company: Spacelabs Healthcare Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Spacelabs Healthcare Inc directly.

Affected Products

Ultraview SL Command Modules, Model 91496, with the Masimo SpO2 PCBA, PN: 010-1636-02. The Spacelabs Multi-parameter Module is intended for use with the Patient Care Management System (PCMS) to acquire, monitor, and process various clinical parameters from an adult or neonate/infant populations in any type of clinical environment other than home use.

Quantity: 1973 units total (1345 modules in the US and 604modules international and 3 PCBAs in US and 21 PCBAs international)

Why Was This Recalled?

Modules resetting. This failure mode happens when the Spacelabs Command Module is configured with the Masimo SpO2 option (-M), the Spacelabs Respiration option (-R) and the Masimo SpO2 PCBA, PN: 010-1136-02. With these two options and the Masimo SpO2 PCBA present, the module may experience random resets.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Spacelabs Healthcare Inc

Spacelabs Healthcare Inc has 19 total recalls tracked by RecallDetector.