Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Spacelabs Healthcare Qube Compact Monitor Recalled by Spacelabs Healthcare Inc Due to Spacelabs Healthcare is recalling Qube Compact Monitor, Model...

Date: June 6, 2014
Company: Spacelabs Healthcare Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Spacelabs Healthcare Inc directly.

Affected Products

Spacelabs Healthcare Qube Compact Monitor, Model 91390. Intended to alert the user to alarm conditions that are reported by Spacelabs Healthcare parameter modules and/or other physiologic monitors via Flexport interfaces.

Quantity: 3383 in the USA; 1459 outside the USA

Why Was This Recalled?

Spacelabs Healthcare is recalling Qube Compact Monitor, Model 91390, due to a potential chance of losing network connection after switching from hardwired Ethernet connection to using the Qube Docking Station without powering down the device. The bedside monitor and all of its alarms will continue to function normally. No one has been injured as a result of this issue.

Where Was This Sold?

This product was distributed to 35 states: AL, AZ, CA, CO, FL, GA, ID, IL, IN, IA, KS, KY, LA, ME, MA, MI, MN, MS, MO, MT, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, WA, WI, WY