Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

ARIATELE TELEMETRY TRANSMITTER Recalled by Spacelabs Healthcare Inc Due to The transmitter display and the Telemetry Central Station...

Date: July 2, 2014
Company: Spacelabs Healthcare Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Spacelabs Healthcare Inc directly.

Affected Products

ARIATELE TELEMETRY TRANSMITTER, Model 96281, with SpO2 Option C. Physiological patient monitor. Used in a hospital environment.

Quantity: 8 units

Why Was This Recalled?

The transmitter display and the Telemetry Central Station display may show a SpO2 value when the sensor is not connected to the patient or in certain sensor failed conditions. This is an expansion of recall Z-0871-2014 because additional serial numbers were identified as being part of this recall action as part of the firm's retrospective analysis of all recalls since 2011.

Where Was This Sold?

This product was distributed to 1 state: TX

Affected (1 state)Not affected

About Spacelabs Healthcare Inc

Spacelabs Healthcare Inc has 19 total recalls tracked by RecallDetector.