Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

S5 Single Roller Pump 150 Recalled by Sorin Group USA, Inc. Due to Sorin Group USA, Inc. has issued an Urgent...

Date: September 19, 2012
Company: Sorin Group USA, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Sorin Group USA, Inc. directly.

Affected Products

S5 Single Roller Pump 150, Item Number: 10-80-00 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.

Quantity: 35 units

Why Was This Recalled?

Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Sorin Group S5 Perfusion System due to reports of a "Fault in motor controller" alarm message displayed on the system panel. This could lead to unintended pump stoppage which may result in permanent impairment or life-threatening injury. An out of specification electrical component was determi

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Sorin Group USA, Inc.

Sorin Group USA, Inc. has 52 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report