Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Coronary Artery Perfusion Cannula with Balloon (part number CP-2200X). The Recalled by Sorin Group USA, Inc. Due to The tip of the cannula should be bent...

Date: August 12, 2016
Company: Sorin Group USA, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Sorin Group USA, Inc. directly.

Affected Products

Coronary Artery Perfusion Cannula with Balloon (part number CP-2200X). The Coronary Artery Perfusion Cannula with Balloon The Cannula consists of a single lumen vinyl tube provided for with a vinyl female luer connector bonded proximally and a vinyl cuff bonded distally. The cuff is pre-molded to its inflated configuration and is purposely positioned directly over a hole generated in the extruded tube for purposes of receiving the infused cardioplegia solution. These devices come in direct contact with the central circulatory system but they are not intended to control, diagnose, monitor or correct a defect.

Quantity: 1205 units worldwide

Why Was This Recalled?

The tip of the cannula should be bent at either a 45 or 90 degree angle. Samples have been returned where the angle of the tip is less than the requirement.

Where Was This Sold?

This product was distributed to 3 states: IL, MO, TX

Affected (3 states)Not affected

About Sorin Group USA, Inc.

Sorin Group USA, Inc. has 52 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report