Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

Medfusion Syringe Pump Model: Model 4000-0100-50 Recalled by Smiths Medical ASD Inc. Due to Multiple issues with the potential for interruption of...

Date: April 18, 2022
Company: Smiths Medical ASD Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Smiths Medical ASD Inc. directly.

Affected Products

Medfusion Syringe Pump Model: Model 4000-0100-50, Model 4000-0101-249, Model 4000-0101-50, Model 4000-0101-51, Model 4000-0101-78, Model 4000-0105-249, Model 4000-0105-51, Model 4000-0105-78, Model 4000-0106-00, Model 4000-0106-01, Model 4000-0106-231

Quantity: 58671 units

Why Was This Recalled?

Multiple issues with the potential for interruption of therapy or over-infusion: 1. Primary Audible Alarm (PAA), 2. Unanticipated Depleted Battery Alarms, 3. Time Base Alarm, 4. Intermittent Volume Over Time (IVOT) - Infusion Continues after System Failure, 5. Clearing of Program Volume Delivered (PVD), 6. False Alarm for Rate Below Recommended Minimum for Syringe Size, 7. Incorrect Bolus or Loading Dose Time Display, 8. Domain Name Server (DNS) Port 1001

Where Was This Sold?

This product was distributed nationwide across all 50 states.