Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Medfusion Model 3500 and 4000 Syringe Infusion Pumps and Medfusion Recalled by Smiths Medical ASD Inc. Due to There have been intermittent issues with Supercap POST...

Date: December 3, 2015
Company: Smiths Medical ASD Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Smiths Medical ASD Inc. directly.

Affected Products

Medfusion Model 3500 and 4000 Syringe Infusion Pumps and Medfusion 3010 and 3010A Syringe Infusion Pumps serviced during the same time period in which a main board was replaced using replacement main board part numbers G6000361, G6000435, G6001260, G6001560, G6001562 and G6002728.

Quantity: 38,797 (31,065 pumps and 7,732 boards)

Why Was This Recalled?

There have been intermittent issues with Supercap POST (power on self-test) and Backup Audible Alarm error codes outside of the pump warranty period. These alarms occur at start up when the pump goes through the self-test. This recall occurred in 2015.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Smiths Medical ASD Inc.

Smiths Medical ASD Inc. has 169 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report