Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

medex TranStar Single OR Set 1/EA Recalled by Smiths Medical ASD, Inc. Due to Due to a manufacturing defect, users may experience...

Date: September 24, 2024
Company: Smiths Medical ASD, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Smiths Medical ASD, Inc. directly.

Affected Products

medex TranStar Single OR Set 1/EA, Product Code REF MX20846R2

Quantity: 260 units

Why Was This Recalled?

Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

Where Was This Sold?

Worldwide distribution.