Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

DuraLife Autoclavable Double Swivel Elbow Recalled by Smiths Medical ASD Inc. Due to DuraLife devices instructions for cleaning and sterilization have...

Date: April 5, 2023
Company: Smiths Medical ASD Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Smiths Medical ASD Inc. directly.

Affected Products

DuraLife Autoclavable Double Swivel Elbow, REF 60-0010 (UltraSet Product Code 66-1991); Swivel Elbow with Suction Port for use with breathing circuits

Quantity: 47007 units

Why Was This Recalled?

DuraLife devices instructions for cleaning and sterilization have been determined to be inadequate. Specifically, for the DuraLife product, the IFU does not provide the cleaning method, duration of autoclave cycle or specify the number of cycles the product can be reused on a single patient.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Smiths Medical ASD Inc.

Smiths Medical ASD Inc. has 169 total recalls tracked by RecallDetector.