Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

CADD-Solis VIP ambulatory infusion pump Recalled by Smiths Medical ASD Inc. Due to Specific Pumps may exhibit intermittent performance in the...

Date: March 24, 2020
Company: Smiths Medical ASD Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Smiths Medical ASD Inc. directly.

Affected Products

CADD-Solis VIP ambulatory infusion pump, Model Numbers: a. 21-2120-0105-02L b. 21-2120-0105-03L c. 21-2120-0105-06L d. 21-2120-0105-07L e. 21-2120-0105-08L f. 21-2120-0105-12L g. 21-2120-0105-13L h. 21-2120-0105-14L i. 21-2120-0105-15L j. 21-2120-0105-17L k. 21-2120-0105-50L l. 21-2127-0105-02L m. 21-2127-0105-06L n. 21-2127-0105-50L

Quantity: 642 units

Why Was This Recalled?

Specific Pumps may exhibit intermittent performance in the AILD (Air in Line Detector) function. If an AILD does not sufficiently discern fluid from air in line, an air in line event may not be recognized by the pump and may not alarm to notify the clinician.

Where Was This Sold?

This product was distributed to 1 state: DE

Affected (1 state)Not affected

About Smiths Medical ASD Inc.

Smiths Medical ASD Inc. has 169 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report