Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Smith & Nephew Taylor Spatial Frame Website software Recalled by Smith & Nephew, Inc. Due to In rare circumstances, when surgeons use the software...

Date: December 13, 2017
Company: Smith & Nephew, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Smith & Nephew, Inc. directly.

Affected Products

Smith & Nephew Taylor Spatial Frame Website software, Catalog Number 71070401. A web-based and mobile application software component of a multilateral external fixation system that is intended for fracture fixation, fixation of long bones and for joint fusions and limb lengthening or deformity corrections which involve cutting of the bone.

Quantity: N/A

Why Was This Recalled?

In rare circumstances, when surgeons use the software to plan the strut adjustment, an inverted image of the frame is generated by the software. The frame will appear inverted in the preview images on the Strut Settings tab.

Where Was This Sold?

Worldwide

About Smith & Nephew, Inc.

Smith & Nephew, Inc. has 220 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report