Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

SL PLUS MIA DOUBLE OFFSET ADAPTER RIGHT 60/25MM Recalled by Smith & Nephew, Inc. Due to The pin that holds the spring mechanism is...

Date: January 7, 2019
Company: Smith & Nephew, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Smith & Nephew, Inc. directly.

Affected Products

SL PLUS MIA DOUBLE OFFSET ADAPTER RIGHT 60/25MM, S&N 75004613

Quantity: 72 units

Why Was This Recalled?

The pin that holds the spring mechanism is not properly mounted. The pin could potentially disassemble.

Where Was This Sold?

This product was distributed to 2 states: NY, TN

Affected (2 states)Not affected

About Smith & Nephew, Inc.

Smith & Nephew, Inc. has 220 total recalls tracked by RecallDetector.