Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

META-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM Recalled by Smith & Nephew, Inc. Due to Integrated Lag/Compression Screw Kit has a compression screw...

Date: March 18, 2025
Company: Smith & Nephew, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Smith & Nephew, Inc. directly.

Affected Products

META-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690

Quantity: 75

Why Was This Recalled?

Integrated Lag/Compression Screw Kit has a compression screw with hexagon socket that may be undersized, which causes the screw to be non-functional, which would require replacement with another adequate size screw, and exchange may exceed 30 minutes.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Smith & Nephew, Inc.

Smith & Nephew, Inc. has 220 total recalls tracked by RecallDetector.