Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

LEGION AP Cutting Block Size 4 Recalled by Smith & Nephew, Inc. Due to Multiple lots of LEGION AP Femoral Cutting Blocks...

Date: December 16, 2019
Company: Smith & Nephew, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Smith & Nephew, Inc. directly.

Affected Products

LEGION AP Cutting Block Size 4, REF 71434409; LEGION AP Cutting Block Size 6, REF 71434411; LEGION AP Cutting Block Size 7, REF 71434412; LEGION AP Cutting Block Size 8, REF 71434413. (Orthopaedic manual surgical instrument).

Quantity: 59 units

Why Was This Recalled?

Multiple lots of LEGION AP Femoral Cutting Blocks were manufactured with an oversize green plunger hole and therefore do not meet their design specifications.

Where Was This Sold?

This product was distributed to 5 states: DE, IA, NC, PA, TN