Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Skytron Integrity 270 Steam Sterilizer and Skytron Integrity 270VP Steam Recalled by Skytron, Div. The KMW Group, Inc Due to Weld attaching the air heater assembly to the...

Date: April 18, 2019
Company: Skytron, Div. The KMW Group, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Skytron, Div. The KMW Group, Inc directly.

Affected Products

Skytron Integrity 270 Steam Sterilizer and Skytron Integrity 270VP Steam Product Usage: The SKYTRON Integrity 270 Steam Sterilizer for use in health care facilities. The Integrity Sterilizer incorporates high-pressure steam to sterilize nonporous or porous items and materials used in the health care industry

Quantity: 5 units

Why Was This Recalled?

Weld attaching the air heater assembly to the main sterilizer unit has shown signs of fatigue/failure

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Skytron, Div. The KMW Group, Inc

Skytron, Div. The KMW Group, Inc has 9 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report