Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

EZ Slide Battery/line powered hydraulic table with removable leg section Recalled by Skytron, Div. The KMW Group, Inc Due to The potential exists within the identified tables that...

Date: April 6, 2016
Company: Skytron, Div. The KMW Group, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Skytron, Div. The KMW Group, Inc directly.

Affected Products

EZ Slide Battery/line powered hydraulic table with removable leg section and 21 top slide General Purpose Surgical Table Skytron EZ Slide Operating Table General Purpose Surgical Table

Quantity: 458

Why Was This Recalled?

The potential exists within the identified tables that incorrectly manufactured parts have been used in assembly. If a table is found to have these incorrectly manufactured parts, it is possible for the tables elevation column slide cap to become detached from the elevation column and could result in the loss of control of the table top and injury to the patient or staff.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Skytron, Div. The KMW Group, Inc

Skytron, Div. The KMW Group, Inc has 9 total recalls tracked by RecallDetector.