Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Ergon III Series Skyboom Monitor Bracket. Part # 3FCM1 Flatscreen Bracket Recalled by Skytron, Div. The KMW Group, Inc Due to Potential for the weld to fail at the...

Date: March 28, 2016
Company: Skytron, Div. The KMW Group, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Skytron, Div. The KMW Group, Inc directly.

Affected Products

Ergon III Series Skyboom Monitor Bracket. Part # 3FCM1 Flatscreen Bracket, single monitor mount, 27inch max. screen size

Quantity: 24

Why Was This Recalled?

Potential for the weld to fail at the pivot point of the monitor bracket. This weld failure could cause the display interface and bracket to fall from its mounted location, resulting in injury.

Where Was This Sold?

This product was distributed to 6 states: GA, IL, NJ, UT, VT, VA

Affected (6 states)Not affected

About Skytron, Div. The KMW Group, Inc

Skytron, Div. The KMW Group, Inc has 9 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report