Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

YSIO X.Pree Recalled by Siemens Medical Solutions USA, Inc Due to For the automated multi-image-acquisition procedure Ortho x-ray collimation...

Date: March 7, 2022
Company: Siemens Medical Solutions USA, Inc
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc directly.

Affected Products

YSIO X.Pree

Quantity: 53 units

Why Was This Recalled?

For the automated multi-image-acquisition procedure Ortho x-ray collimation is set in a preparative stage for the entire examination area prior to the exam. During acquisition of each individual x-ray image, the x-ray collimator is automatically positioned in a way that the subsequent series of acquisition covers the defined field of view needed for each step. However, during the acquisition the collimation area displayed to the operator on the User Interface does not correctly represent the collimation area specified by the system. It indicates to the user an open collimator instead, e.g., the information displayed on the User Interface shows wider area of collimation than values preset prior to the examination. However, the collimation of the x-ray is performed correctly and always matches the examination area predefined by the user.

Where Was This Sold?

This product was distributed nationwide across all 50 states.