Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Ysio MAX (model number 10762470) with Tube stand 3D V Recalled by Siemens Medical Solutions USA, Inc Due to In rare cases of high clinical workload, the...

Date: August 6, 2018
Company: Siemens Medical Solutions USA, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc directly.

Affected Products

Ysio MAX (model number 10762470) with Tube stand 3D V (model number 7042018 or 7042034), radiographic systems used in radiographic and tomographic exposures of the whole body.

Quantity: 31

Why Was This Recalled?

In rare cases of high clinical workload, the steel cables inside the lifting column of the overhead tube, which are designed to take the load, may rapture without triggering the safety lock. This could lead to overloading the safety rope, which is not designed for continuous load and ongoing movement under load. This may lead to mechanical fatigue and cause the arm to drop down unexpectedly during movement in vertical direction and during patient positioning.

Where Was This Sold?

This product was distributed to 37 states: AL, AZ, AR, CA, CO, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, MA, MI, MN, MS, MO, NE, NV, NM, NY, NC, ND, OH, OK, PA, SC, SD, TX, UT, VA, WI, WY, DC

Affected (37 states)Not affected

About Siemens Medical Solutions USA, Inc

Siemens Medical Solutions USA, Inc has 613 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report