Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Torque wrench used on the bolts for the Symbia T2 model 10275008 Recalled by Siemens Medical Solutions USA, Inc. Due to The torque wrench used to tighten system bolts...

Date: September 18, 2017
Company: Siemens Medical Solutions USA, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc. directly.

Affected Products

Torque wrench used on the bolts for the Symbia T2 model 10275008, Symbia Intevo Excel model 10764801, Symbia Intevo 16 model 10764804, Symbia Evo model 10910500, and Symbia Evo Excel model 10910501 devices.

Quantity: 16 devices

Why Was This Recalled?

The torque wrench used to tighten system bolts during installation was found to be out of tolerance

Where Was This Sold?

This product was distributed to 7 states: CA, FL, GA, LA, NY, PA, TN