Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
The Symbia Intevo series consists of the Intevo 16 Recalled by Siemens Medical Solutions USA, Inc. Due to Siemens Medical Solutions USA, Inc. is performing a...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc. directly.
Affected Products
The Symbia Intevo series consists of the Intevo 16, Intevo 6 and Intevo 2. This manual also describes the Symbia Intevo Excel. The Symbia Intevo series integrates state-of-the-art SPECT and high quality spiral CT to provide precise attenuation correction and anatomical mapping. The system can also be used as a complete clinical SPECT system for general purpose, whole body (WB) and SPECT applications. The Symbia Intevo Excel integrates state-of-the-art SPECT and high quality spiral CT to give the system SPECT functionality, with attenuation correction. The Symbia Intevo 2 integrates state-of-the-art SPECT and high quality dual slice spiral CT to give the system full functionality for all SPECT-only, xSPECT, or stand-alone CT diagnostic applications in Oncology, Neurology, and Cardiology. The Symbia Intevo 6 integrates state-of-the-art SPECT and high quality six slice spiral CT to give the system full functionality for all SPECT-only, xSPECT, or stand-alone CT diagnostic applications in Oncology, Neurology, and Cardiology. The Symbia Intevo 16 integrates state-of-the-art SPECT and high quality sixteen slice spiral CT to give the system full functionality for all SPECT-only, xSPECT, or stand-alone CT diagnostic applications in Oncology, Neurology, and Cardiology. Indications for use: SPECT: To detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588keV. CT: The CT component is intended to produce cross sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial place taken at different angles or spiral planes taken at different angles. SPECT+CT: Perform CT scans and nuclear imaging studies within the same instrument. To obtain attenuation corrected images and to provide registration of anatomical and physiological images within the patient's anatomy.
Quantity: 48 systems
Why Was This Recalled?
Siemens Medical Solutions USA, Inc. is performing a field correction because we recently detected a rubber isolator with unfavorable aging properties. The function of the rubber isolator, which is situated within the rotate motor mounting assembly, may degrade over time and may impact motor support. The event does not pose a hazard by itself but could lead to a rupture of the drive belt which in t
Where Was This Sold?
This product was distributed to 38 states: AL, AK, AZ, AR, CA, CO, CT, FL, GA, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MO, NV, NH, NJ, NM, NY, NC, OH, OR, PA, SC, TN, TX, UT, VT, VA, WI, WY
About Siemens Medical Solutions USA, Inc.
Siemens Medical Solutions USA, Inc. has 87 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report