Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

The Artis zee / zeego Angiography System is designed as Recalled by Siemens Medical Solutions USA, Inc Due to There is a potential issue on running Artis...

Date: September 11, 2013
Company: Siemens Medical Solutions USA, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc directly.

Affected Products

The Artis zee / zeego Angiography System is designed as a set of components that may be combined into different configurations to provide specialized angiography systems.

Quantity: 1131

Why Was This Recalled?

There is a potential issue on running Artis systems running software VC1x software if a network problem arises, the function cannot be deactivated again by pressing the "Block Radiation" key on the touch screen control. If this behavior occurs, the system image generation function is not available for patient examinations without switching the system off and back on again manually. This

Where Was This Sold?

This product was distributed nationwide across all 50 states.