Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Syngo.Via VB20A model 1049610 Recalled by Siemens Medical Solutions USA, Inc Due to Functionality in the report sections "Findings Information" and...

Date: March 27, 2018
Company: Siemens Medical Solutions USA, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc directly.

Affected Products

Syngo.Via VB20A model 1049610

Quantity: N/A

Why Was This Recalled?

Functionality in the report sections "Findings Information" and "Summary of Measured Findings" not functioning properly. The corrected values modified in the report are neither saved, nor printed, nor sent to the information system. The system will still use original values.

Where Was This Sold?

This product was distributed nationwide across all 50 states.