Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Syngo.via software is intended to be used for viewing Recalled by Siemens Medical Solutions USA, Inc Due to When the archiving configuration is changed, data received/created...

Date: January 10, 2018
Company: Siemens Medical Solutions USA, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc directly.

Affected Products

Syngo.via software is intended to be used for viewing, manipulation, communication, and storage of medical images. It can be used as a standalone device or together with a variety of cleared and unmodified syngo based software options.

Quantity: 165 units

Why Was This Recalled?

When the archiving configuration is changed, data received/created after the upgrade may be flagged as "Not to be archived". This is caused by the automatic function for cleaning up temporary data being disabled by the software upgrade. Due to the disabled cleanup function, disc capacity for free space decreases faster than usual. Unless the archiving configuration and the cleanup automatic function are reverted back to the original state (prior to the software upgrade), data that was incorrectly flagged "Not to be archived" must be manually prevented from being deleted by either clinical administrator or service engineer.

Where Was This Sold?

Medical device software which needs to be installed.

About Siemens Medical Solutions USA, Inc

Siemens Medical Solutions USA, Inc has 613 total recalls tracked by RecallDetector.