Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

syngo.via RT Image Suite upgraded to VB60 running on SOMATOM Recalled by Siemens Medical Solutions USA, Inc Due to Upgraded software version makes an automatic change in...

Date: March 7, 2022
Company: Siemens Medical Solutions USA, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc directly.

Affected Products

syngo.via RT Image Suite upgraded to VB60 running on SOMATOM go.Open Pro Model Number 11061670. Running on servers with syngo.via VB60 software after upgrade and CT scanners with syngo.CT VA40A after upgrade.

Quantity: 14 (US)

Why Was This Recalled?

Upgraded software version makes an automatic change in laser configuration settings for "markerless workflow" which may lead to the wrong laser offset coordinates being displayed in the Patient Marking step.