Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

syngo Dynamics Version 9.5 Recalled by Siemens Medical Solutions, USA, Inc Due to If multiple uncompressed images/clips are sent to the...

Date: July 13, 2012
Company: Siemens Medical Solutions, USA, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions, USA, Inc directly.

Affected Products

syngo Dynamics Version 9.5

Quantity: 30

Why Was This Recalled?

If multiple uncompressed images/clips are sent to the syngo Dynamics server, around the same time, it is possible for an image from a patient to be placed in another patient's study.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Siemens Medical Solutions, USA, Inc

Siemens Medical Solutions, USA, Inc has 1 total recall tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report