Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Syngo Dynamics Recalled by Siemens Medical Solutions USA, Inc Due to Siemens has identified changes in default carry forward...

Date: April 14, 2016
Company: Siemens Medical Solutions USA, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc directly.

Affected Products

Syngo Dynamics; Kinetdx Picture Archiving and Communications System

Quantity: 84 units

Why Was This Recalled?

Siemens has identified changes in default carry forward workflow using VA10B_HF03 version of Syngo Dynamics. If the template design has measurements inside a carry forward-configured report section, this will result in carrying forward the entire section of the report including data from previous studies. This may result in the physician making analysis based on the old measurements.

Where Was This Sold?

This product was distributed nationwide across all 50 states.