Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

syngo Application software VE20 (Material Number 10848815) installed on the Recalled by Siemens Medical Solutions USA, Inc Due to After CT image data from Toshiba is loaded,...

Date: February 25, 2022
Company: Siemens Medical Solutions USA, Inc
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc directly.

Affected Products

syngo Application software VE20 (Material Number 10848815) installed on the Artis pheno and Artis icono systems as follows: (1) Artis pheno Model Number:10849000 (2) Artis icono biplane Model Number: 11327600 (3) Artis icono floor Model Number: 11327700

Quantity: 127 units

Why Was This Recalled?

After CT image data from Toshiba is loaded, image mirroring can occur along the horizontal and vertical image axes. If this error occurs, the patient orientation/position may be misinterpreted and result in inappropriate treatment, even if the incorrect visualization is obvious.

Where Was This Sold?

This product was distributed nationwide across all 50 states.