Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

SOMATOM Definition AS Recalled by Siemens Medical Solutions USA, Inc Due to Siemens became aware that the front cover on...

Date: March 22, 2016
Company: Siemens Medical Solutions USA, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc directly.

Affected Products

SOMATOM Definition AS, a family of CT systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission dta from either the same axial plane taken at different angles or spiral planes taken at different angles.

Quantity: 1 CT system

Why Was This Recalled?

Siemens became aware that the front cover on some Siemens CT system was not properly welded. It is possible for the welded-in threaded bolt to break out of the ground plate of the front cover if the cover is opened during service. This could not happen during clinical operation. Therefore, there is no potential risk for the user or patient and the CT System has no risk for a malfunction during clinical operation